| Thermometry |
| Number of Inputs |
Two included (additional inputs optional) |
| Measurement Type |
4-lead Differential |
| Sensor Types (included) |
Silicon Diode, GaAIAs Diode, Platinum RTD, Cernox RTD,
Germanium RTD, Carbon-Glass RTD, Ruthenium Oxide, Rhodium Iron, other resistive sensors |
| Sensor Types (Optional) |
Thermocouple, Capacitance |
| Input Accuracy |
Sensor dependent |
| A/D Resolution |
24-bit analog-to-digital |
| Update Rate |
Up to 20 readings/second on an input, 40 readings/second
on all inputs |
| User Curve Storage |
Forty, 200-point CalCurves, or user curves |
| SoftCal |
Improves accuracy of DT-470 diode or platinum RTD sensors |
| Math |
Maximum and minimum of input readings and linear equation |
| Filter |
Averages input readings to quiet display, settable time
constant |
| |
|
Control |
| Control Loops |
Two |
| Control Type |
Digital PID with manual output |
| Control Stability |
Sensor dependent |
| |
|
Heater
Output (Loop 1 Only) |
| Heater Output Type |
Variable DC current source |
| Heater Output Resolution |
18-bit digital-to-analog converter |
| Max Heater Power |
100 W |
| Max Heater Output Current |
2A |
| Current Limit Settings |
2 A, 1 A, 0.5 A, 0.25 A |
| Heater Output Compliance |
50 V |
| Heater Output Ranges |
5 decade steps in power |
| |
|
Heater Load |
| Heater Load Type |
Resistive |
| Heater Load Range |
10 W to 100 W, 25 W for max heater power |
| Heater Noise |
50 mV + 0.001% of output voltage |
| |
|
PID Control Settings |
| Gain (Proportional) |
0-1000 with 0.1 setting resolution |
| Reset (Integral) |
1-1000 with 0.1 setting resolution |
| Rate (Derivative) |
1-1000 seconds with 1 second resolution |
| Manual Output |
0-100% with 0.01% setting resolution |
| Zone Control Mode |
10 temperature zones with control parameters |
| Ramping |
Setpoint ramping at user specified rate |
| Internal Program |
100 program steps |
| |
|
Front Panel |
| Display Type |
Graphic LCD with fluorescent backlight |
| Number of Input Displays |
One to eight |
| Display units |
Temperature in K, °C, or sensor units |
| Display Resolution Input Temperature |
0.0001 K below 10 K, 0.001 K above 10 K |
| Display Resolution Sensor Units |
Sensor dependent, to six digits |
| Setpoint Setting Resolution |
Same as display resolution (actual resolution
is sensor dependent) |
| Heater Output Display |
Numeric display in percent of full scale for
power or current. Bar graph display of Heater Output available |
| Keypad |
Numeric plus special function |
| Curve Entry |
Front panel curve entry |
| |
|
Interfaces |
| IEEE-488.2 Interface |
SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1 |
| Serial Interface |
RS-232C electrical format, 19,200 baud, RJ-11 connector |
| Analog Outputs |
Temperature monitor or Loop 2 control output |
| Number |
Two |
| Voltage Range |
+ 10 volts |
| Resolution |
1.25 mV |
| Maximum Current |
0.1 amperes |
| Minimum Load |
100 W |
| Alarm Type |
High and low alarm for each input |
| Alarm Actuators |
Audible beeper or relays |
| Relays |
|
| Number |
Two |
| Contacts Available |
Normally open (N.O.), normally closed (N.C.), and common |
| Contact Rating |
30 VDC at 2 A |
| Digital I/O |
5 inputs and 5 outputs. TTL voltage level compatible |
| Data Card |
PC card Type II slot used for curve transfer and setup
storage |
| |
|
| General |
| Ambient Temperature Range |
20-30°C (68-86°F) for specified accuracy; 15-35°C (59-95°F) for reduced accuracy |
| Power Requirements |
100, 120, 220, 240 VAC (+ 5%, -10%), 50 or 60
Hz; 200 VA |
| Enclosure Type |
Full 19-inch rack mount |
| Size |
431.8 mm x 88.9 mm x 381 mm (17" x 3.5" x
15") |
| Weight |
11.4 kilograms (25 pounds) |
| |
|