DMX-5 Specifications
Cooling Technology- | ||
---|---|---|
DE-202 | Closed Cycle Cryocooler |
|
Refrigeration Type | Pneumatically Driven Gifford-McMahon Cycle | |
Liquid Cryogen Usage | None, Cryogen Free | |
Temperature*- | ||
DE-202AE | < 10 K - 350 K | |
DE-202SE | < 4.2 K - 350 K | |
DE-202PE | < 5.5 K - 350 K | |
With 800 K Interface | (Base Temp + 2 K) - 700 K | |
With 450 K Interface | (Base Temp + 2 K) - 450 K | |
Stability | 0.1 K | |
*Based on bare cold head with a closed radiation shield, and no additional sources of experimental or parasitic heat load. | ||
Sample space- | ||
Diameter | 36 mm (1.43 in) | |
Height | 39 mm (1.53 in) | |
Sample Holder Attachment | 1/4-28 screw | |
Sample Holder | View our Sample Holder Collection | |
Optical Access- | ||
Window Ports | 360° viewing angle | |
Diameter | 4 inches long | |
Clear view | N/A | |
#/F | N/A | |
Window Material | View our Wide Selection of Window Materials | |
Temperature Instrumentation and Control (Standard)- | ||
Heater | 1 - 36ohm Thermofoil Heater Anchored on Cold Tip | |
Control Sensor | 1 - Curve Matched Silicon Diode | |
Sample Sensor | 1 - Calibrated Silicon Diode | |
Custom Instrumentation | Contact ARS for available Options | |
Instrumentation Access- | ||
Instrumentation Skirt | Bolt-on, Aluminum | |
Pump out Port | 1 - NW-25 | |
Instrumentation Ports | 2 | |
Instrumentation Wiring | Contact our sales staff for wiring options | |
Vacuum Shroud- | ||
Material | Aluminum | |
Length | 397 mm (15.62 in) | |
Diameter | 76 mm (2.75 in) (at the sample space) | |
Width | 76 mm (2.75 in) (at the sample space) | |
Radiation Shield- | ||
Material | Aluminum | |
Attachment | Threaded | |
Optical Access | 0, 2, or 4(customer specified) | |
Cryostat Footprint- | ||
Overall Length | 619 mm (24.34 in) | |
Motor Housing Diameter | 114 mm (4.5 in) | |
Rotational clearance | 200 mm (8 in) with "G" configuration |